Error Code E-10: Checksum failure (the program’s code has a problem)
Error Code E-11: RAM failure (the problem’s memory/space has a problem)
Error Code E-12: Switch failure (the machine’s switches have short circuited)
Error Code E-13: Time failure (the machine’s time does not align with the on-board crystal)
Error Code E-20: Open probe (check for broken leads)
Error Code E-21: Reasonableness failure (check for any erratic temperature readings)
Error Code E-30: Over heating error
Error Code E-31: Heater burned out and machine emitting little or no heat (possibly need to replace that part)
Note: Error codes E-30 and E-31 both can be from dust buildup in the back panel or broken fans
Error Code E-40: NVRAM read failure (need to replace control board)
Error Code E-41: NVRAM write failure (need to replace control board)