{"id":728,"date":"2017-08-11T13:35:17","date_gmt":"2017-08-11T13:35:17","guid":{"rendered":"https:\/\/learn.sterilizers.com\/?p=728"},"modified":"2017-08-14T04:03:27","modified_gmt":"2017-08-14T04:03:27","slug":"cox-dry-heat-sterilizer-cylinder-filled-water-cycle","status":"publish","type":"post","link":"https:\/\/learn.sterilizers.com\/blog\/cox-dry-heat-sterilizer-cylinder-filled-water-cycle\/","title":{"rendered":"Pelton Crane – Chamber Filled with Water After Cycle is complete"},"content":{"rendered":"

This can be caused by either bad fill solenoid valve, or debris in the piping system of the fill line.<\/p>\n

Run the sterilizer up to the operating pressure and temperature.
\nTurn the bottom knob to ‘fill position’.
\nGive it a minute or two for venting to complete.
\nThis should blow out any debris in the line and solenoid.
\nIf after doing this you still have the problem, replace the Solenoid<\/span><\/strong>.<\/a><\/p>\n

Thank you –
\n~Shlomo Savyon
\n1-800-762-1586
\n
Your Sterilizer expert<\/a><\/strong><\/p>\n","protected":false},"excerpt":{"rendered":"

This can be caused by either bad fill solenoid valve, or debris in the piping system of the fill line. Run the sterilizer up to the operating pressure and temperature. Turn the bottom knob to ‘fill position’. Give it a minute or two for venting to complete. This should blow out any debris in the […]<\/p>\n","protected":false},"author":1,"featured_media":0,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"footnotes":""},"categories":[1,63],"tags":[],"_links":{"self":[{"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/posts\/728"}],"collection":[{"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/comments?post=728"}],"version-history":[{"count":7,"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/posts\/728\/revisions"}],"predecessor-version":[{"id":747,"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/posts\/728\/revisions\/747"}],"wp:attachment":[{"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/media?parent=728"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/categories?post=728"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/learn.sterilizers.com\/wp-json\/wp\/v2\/tags?post=728"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}