Q:  How do we fix the TOMY SX-500 Sterilizer when we get error code E09 (Lid Sensor Defective)?

A:  Most likely it’s the right hook microswitch that is not working properly.  Attached is the process of removing the panel and readjusting the microswitch.

Maintenace mode manual C08 DME10D311 SX series Troubleshooting for Er.E09 (1)

DME10D311 SX series Troubleshooting for Er.E09 (1)

Thank you –
~Shlomo Savyon
Your Sterilizer expert

TOMY ES-315 Sterilizer – ERR message – What shall I do?

ERR normally shows a number after indicating the input error (eg. Water level sensor is 1).  Err with no number indicates multiple errors.   If you press start when there is no water in the chamber and the lid is open, you will get an ERR 0 (or ERR according to TOMY).  I have only one case of ERR only and that was related to an external incident that damaged the autoclave severely.  I have never seen this error otherwise (20+ year history).

TOMY changed the circuit board for the new sterilizers around the year 2010, and most likely the ERR without number was added to the logic.

The best course is to make sure the chamber has sufficient water, close the lid and the exhaust valve.  That should isolate the error if there is still one.  If not, they will need to go through the inputs to see which components are failing.  If the inputs test, then the problem is probably with the board.

Thank you,
Shlomo
Sterilizers.com

IQ OQ PQ – What is it?
It’s a factory documentation package which is available upon request when you order a machine with IQ/OQ/PQ

IQ – Preliminary Installation Qualifications

OQ – Preliminary Operation Qualifications

PQ – Preliminary Performance Qualifications

These documents are produces and prepared during the production and the testing.

There is a fee for that.  Each manufacturer has different fee for each of their products.

 

Alfa Medical of Westbury NY recently (January 2015) became approved distributor of Priorclave Sterilizers.

Size choices range from 35 liters to 125 liters for these bench top and top loading models. All equipment offers a thermal safety lock, free steaming, and a microprocessor based TACTROL control system. These simple, slimmed down autoclaves are ideal for sterilizing your laboratory waste and preparing growth media. The pharmaceutical industry is the primary market for these high value – low cost machines. 

Discover more here Priorclave Sterilizers

PriorClave Understands How To Get Things Done Right

This company has been in the autoclave and steam sterilizer business for over two decades and is currently based in London at a 2500 square meter manufacturing facility. PriorClave directs a substantial portion of its profits into continual improvement of their research and manufacturing processes. One of their goals is to bring more of their component creation in-house. Having fewer outside sources involved gives the corporation the ability to reach even higher levels of quality control over every item that goes into their machines.

They currently operate under BS EN ISO9001 requirements with additional oversight of processes and designs provided by Zurich Insurance. PriorClave also makes good use of computer controlled (CNC) equipment to reduce the risk of human error in many of its manufacturing stages.

Safety locks are, of course, standard on all models. These prevent a unit from being opened before the glassware inside is properly cooled (sudden changes in temperature can cause glass to explode). As the icing on the cake, each machine features anti-microbial BioCote on its surfaces to offer continuing decontamination even when the machine is not in use.

For more information, contact
Shlomo Savyon
Your Autoclave & Sterilizer Expert
516-283-5535